Size Range: |
0.5 to 220 microns |
Accuracy: |
±2% of x(50) using approved traceable standards |
Supply voltage: |
+12V DC |
Operating Current: |
350mA |
PC to Sizer Serial Communications: |
RS422 |
PC to PLC |
Serial data & Ethernet to allow output of user selectable result. |
Sampling rate: |
2000 samples per second. |
Sampling resolution: |
16-bit analogue to digital converter. |
Laser: |
5mW 635nm semiconductor laser diode. |
Analogue Output : |
2 x opto isolated 16-bit analogue outputs 0 to 10V & 4 to 20mA |
Digital Inputs: |
2 x general purpose opto isolated digital. |
Digital Outputs: |
3 x general purpose opto isolated open collector digital outputs. |
Local Display: |
2 lines x 16 characters backlit LCD. |
Temperature Range: |
-10 to 50°C standard range:(extended range -30 to + 70°C) |
Industrial protection: |
IP65 as standard |
Maximum distance from instrument to PC: |
1.2km |
Dimensions: |
425 X 95 X 95mm |